Waters ACQUITY RDa Detector
Sign in to view price
$ZGK0APERFORMANCE
Mass Accuracy
RMS error <2.5 ppm (measured using consecutive repeat measurements of sodium and rubidium iodide, with Lock Mass correction);
Mass scale calibration accuracy: <1 ppm with internal Lock Mass correction over range 50 to 2,000 m/z
Resolution
10,000 FWHM (measured on a singly charged ion between 550 and 650 m/z)
Sensitivity
Positive ion MS sensitivity: ≥22,000 ions/counts per second (m/z 556.28);
Negative ion MS sensitivity: ≥13,750 ions/counts per second (m/z 554.26)
Speed
Scan speed: 1 to 20 Hz
Dynamic Range
Chromatographic dynamic range: ≥4 orders of magnitude (measured using UPLC injections of lidocaine, monitoring the ion at m/z 235)