Thermofisher Orbitrap Exploris 480 Mass Spectrometer
Orbitrap Exploris 480: High-resolution mass spectrometer for biopharma, proteomics, and isotope ratio analysis.
Features
- Market-leading resolution & mass accuracy
- Compact footprint
- Isotope ratio analysis with high-resolution accuracy
- Maximum data certainty and robustness
- High throughput for large-scale studies
Applications
- Proteomics
- Biopharma
- Isotope ratio analysis
- Large-scale studies
- Translational science research
Product Specifications
PERFORMANCE
Mass Accuracy
<3 ppm RMS with external calibration over 24 hr. and without any kind of mass recalibration, <1 ppm RMS drift for at least 5 days
Resolution
Up to 480,000 at m/z 2001
Sensitivity
MS/MS: 50 fg Reserpine on Column S/N 100:1
SIM: 50 fg Reserpine on Column S/N 150:1
SIM: 50 fg Reserpine on Column S/N 150:1
Speed
Up to 40 Hz at resolution setting of 7500 at m/z 2002
Scan rate: Up to 40 Hz at resolution setting at 7,500
Scan rate: Up to 40 Hz at resolution setting at 7,500
Dynamic Range
>5000:1 within one spectrum
>5,000 within a single Orbitrap mass analyzer spectrum
>5,000 within a single Orbitrap mass analyzer spectrum
MASS RANGE AND ANALYTE DETECTION
Mass Range
40 to 8000 m/z
Standard mass range m/z 40–6,000
Up to m/z 8,000 with BioPharma option
Standard mass range m/z 40–6,000
Up to m/z 8,000 with BioPharma option
Detection Limit
- Full Scan: < 1.5 fg reserpine on column
- tMS2: < 1.5 fg reserpine on column
- tSIM: < 0.8 fg reserpine on column
Detector
Independent Charge Detector
IONIZATION, SOURCE AND INTERFACE
Ionization Method
Soft electrospray ionization (ESI)
Heated electrospray ionization (H-ESI)
Heated electrospray ionization (H-ESI)
Ion Source
Thermo Scientific™ OptaMax™ NG ion source
EASY-IC™ ion source
EASY-IC™ ion source
Interface
Atmospheric pressure ionization (API) interface
Gas Flow and Requirements
- Source gas: high-purity nitrogen (>99% pure at 600 ± 50 kPa [6.0 ± 0.5 bar, 87 ± 7 psi])
- HCD collision gas: ultra-high-purity nitrogen (UHP > 99.999% pure) at 600 ± 50 kPa [6.0 ± 0.5 bar]
- For EASY IC option: ultra-high-purity nitrogen (UHP > 99.999% pure) at 600 ± 50 kPa [6.0 ± 0.5 bar]
MASS ANALYZER
Mass Analyzer
Orbitrap
Ultra-high-field Thermo Scientific Orbitrap mass analyzer with 4 kV central electrode voltage
Thermo Scientific™ Orbitrap™ mass analyzer
Ultra-high-field Thermo Scientific Orbitrap mass analyzer with 4 kV central electrode voltage
Thermo Scientific™ Orbitrap™ mass analyzer
Quadrupole
Advanced quadrupole technology (AQT)
Segmented quadrupole mass filter for precursor ion selection with variable precursor isolation width from 0.4 to 2,000 Da
Segmented quadrupole mass filter for precursor ion selection with variable precursor isolation width from 0.4 to 2,000 Da
CHROMATOGRAPHY AND SAMPLE HANDLING
Chromatography
LC-MS
Sample Handling
Automatic sample introduction, capable of handling liquid samples via direct infusion or LC systems
Injection System
Compatible with autosamplers or manual injection, with flow rates up to 1 mL/min
DATA ACQUISITION AND ANALYSIS
Data Acquisition
- High-performance PC with Intel® processor
- High-resolution LED monitor
- Windows® Enterprise OS (LTSC)
- Real-time data acquisition & control
- Auto calibration of ion transfer & analysis
Data Analysis
Thermo Scientific™ Orbitrap Exploris™ instrument control software
Thermo Scientific™ Xcalibur™ software
Thermo Scientific™ Xcalibur™ software
AUTOMATION AND SOFTWARE
Automation
AcquireX data acquisition workflows provide comprehensive small molecule sample profiling on an LC timescale using automated, logic-based sample re-injection with inter-run inclusion and exclusion lists
Software
- Thermo Scientific™ Orbitrap Exploris™ software
- Thermo Scientific™ Xcalibur™ software
- Thermo Scientific™ Proteome Discoverer™ software
- Thermo Scientific™ BioPharma Finder™ software
- Thermo Scientific™ ProSightPD™ software
- Thermo Scientific™ Compound Discoverer™ software
- Thermo Scientific™ TraceFinder™ software
- Thermo Scientific™ Chromeleon™ Chromatography Data System (CDS)
OPERATIONAL FEATURES
Ease of Use
Next-generation software with intuitive tuning, single-click calibration, drag-and-drop method setup, and pre-defined experiments
Maintenance and Durability
Prolonged robustness and improved serviceability increase uptime
Operating Conditions
- Operating temperature: 18–27 °C (64–81 °F)
- Relative humidity: 20–80% (no condensation)
- Indoor use up to 3,000 m (10,000 ft) altitude
Power and Cooling
- 2×208–240 Vac single phase, 15 A, 50/60 Hz for instrument and source vacuum pump
- 115 or 230 Vac single phase, 15 A, 50/60 Hz for auxiliary forepump
- 208–240 Vac single phase, 15 A, 50/60 Hz for data system
- System power output: 3440 W (11,730 Btu/h) for air conditioning needs
Dimensions and Footprint
534 × 763 × 703 mm (21 × 30 × 27.7 in)
120 kg (265 lb) without data system, vacuum rough pumps, and optional items
120 kg (265 lb) without data system, vacuum rough pumps, and optional items
GENERAL INFORMATION
Brand and Model
Thermo Scientific™, Orbitrap Exploris™ 480 Mass Spectrometer
Orbitrap Exploris 480 MS
Orbitrap Exploris 480 MS
Documentation and Support
Comprehensive user manuals, application notes, and 24/7 technical support through Thermo Fisher service network
Compatibility and Connectivity
Compatible with Thermo Scientific chromatography systems (e.g., Dionex) and third-party systems via USB, Ethernet, and Wi-Fi
Configuration and Options
- Thermo Scientific™ BioPharma option
- Thermo Scientific™ EASY-Spray™ NG ion source
- Thermo Scientific™ NanoSpray Flex™ NG ion source
- APCI probe for OptaMax NG ion source
- Thermo Scientific™ FAIMS Pro™ Duo interface
Licensing and Certification
For Research Use Only. Not for use in diagnostic procedures
Cost and Operational Expenses
Initial cost: $200,000 - $400,000 USD
Operational expenses include consumables (e.g., columns, reagents), maintenance, and service contracts (~$20,000 per year)
Operational expenses include consumables (e.g., columns, reagents), maintenance, and service contracts (~$20,000 per year)